Recommended Maintenance Schedule

Climatronics recommends that system maintenance begins by establishing a schedule for periodic system reviews. These reviews should be conducted monthly and semi-annually. It may be useful after the system is first installed, to do some of the checks more frequently until the system operation is completely familiar.

Monthly site checks are reviews of the system to make sure that it is reporting reasonable data. Should this check indicate that there is a problem with the system, corrective maintenance should be performed.

Periodic maintenance is work to be performed every six months or more frequently should your operational experience indicate the need. This should keep your system operational and prevent non-scheduled outages.

Monthly Site Checks
     
100075
Wind Speed Sensor
Check to see that the cups are turning.
100076/102139
Wind Direction Sensor
Check to see that the data appear reasonable.
100093
Temperature Sensor
Compare data to psychrometer.
102273/101943
Relative Humidity Sensor
Compare data to psychrometer. Check sensor filter for dirt/dust. Clean or replace as necessary.
100325
Motor-Aspirated Shield
Clean front inlet and rear housing.
102318/102319
Pyranometer
The outer dome must be cleaned and inspected. Check it to ensure that the instrument is level and that there is no condensation inside the dome.
100097/100508
Precipitation Gauge
Check funnel and screen, and remove any debris.

Semi-Annual Periodic Maintenance
     
100075
Wind Speed Sensor
Replace sensor on tower with spare. Use wind speed test fixture to check speed, comparing both the digital and analog output. Replace bearings on sensor removed from tower.
100076/102139
Wind Direction Sensor
Replace sensor on tower with spare. Replace bearings on sensor removed from tower. Use linearity fixture when re-assembling the sensor.
100093
Temperature Sensor
Remove sensor from aspirator and wipe any dust or dirt from sensor with a clean cloth. Replace in aspirator when finished.
102273/101943
Relative Humidity Sensor
Remove sensor from aspirator. Look at sensor to be sure it is clean. Compare data to psychrometer. Recalibrate if necessary.
100325-5
Motor-Aspirated Shield
Clean aspirator of any dirt or dust. Feel fan exhaust to make sure fan is running.
102318/102319
Pyranometer
Replace the desiccant twice a year in humid areas, or as required in other locations.
100097/100508
Precipitation Gauge
Removal funnel. Clean funnel, bucket assembly, drain tubes, drain line, and inner funnel. Oil pivot bearing with a drop of light oil. Check the calibration per the manual's calibration procedure.

 

 

 


Service Bulletins with
Spare Parts Pricing:


101908 WM III Speed & Direction Sensor 4-20 Current Output (92k)

102083-100108 WM III Speed & Direction Sensor (90k)

100075 F460 Wind Speed Sensor (91k)

100076 F460 Wind Direction Sensor (18 K)

102139 F460 Resolver Wind Direction Sensor (91k)

102236 Vertical Wind Speed Sensor (16k)

100093-102090 Temperature Sensor (13 K)

100093-2-102096 Temperature Sensor (14 K)

102304 TACMET II Weather Station (16K)

100097 8 In. Precipitation Gauge (16k)

100508 6 In. Precipitatipn Gauge(16k)

102663 Barometic Pressure (16k)

102766 TACMET II Weather Sensor

102779 SONIC Anemometer

102780 AIO (ALL IN ONE) Wind Sensor with AT, RH, BP

102795 TACMET II